Site 02, PA Gamelands Instrument Log

Site 02, PA Gamelands, susq06, CFDBS2042, Flux instrument (CO2/CH4/H2O)
41.9063, -75.6935
measurement height = 46 m AGL

May 24, 2013: Installed in tower building.  Flux pump with flow restrictor (part # A-18-NY).  Tested tank CC336384 (NOAA values:  395.76 ppm CO2 / 143.56 ppb CO / 1865.29 ppb CH4).  Measured values:  394.35 ppm CO2 / 1867.7 ppb CH4.

DID NOT LEAVE MOUSE.  KEY MUST BE RETURNED AT END OF PROJECT.
Tower climbers secured tubing to tower with electrical tape.
CALL PRIOR TO ENTERING.

July 30, 2013:  Decommissioned site.  NOAA:  395.76 ppm CO2, 143.56 ppb CO, 1865.29 ppb CH4.  Measured:  393.85 ppm CO2,1863.5 ppb CH4.  NOAA:  451.17 ppm CO2, 254.49 ppb CO, 2117.14 ppb CH4.  Measured:  449.25 ppm CO2, 2116.5 ppb CH4.
measured – NOAA

Site 02

Pre: -1.41ppm CO2, Post(Low tank/High tank):  -1.9 ppm/ -1.9 ppm, Post-lab(Low tank/High tank): -1.8 ppm / -1.8 ppm

Pre: 2.4 ppb CH4, Post(Low tank/High tank):  -1.8 ppb / -0.6 ppb, Post-lab(Low tank/High tank):  -0.6 ppb / 1 ppb

**When having tower climbed to remove tubing, ask for funnels and antenna to be returned.

 